TEMPERATURE AND HUMIDITY CHAMBER
1. The main technology parameter:
model |
THP80 |
THP100 |
THP150 |
THP225 |
THP408 |
THP800 |
THP1000 |
volume |
80L |
100L |
150L |
225L |
408L |
800L |
1000L |
D400 |
D400 |
D500 |
D500 |
D600 |
D800 |
D1000 |
|
W400 |
W500 |
W500 |
W600 |
W800 |
W1000 |
W1000 |
|
H500 |
H500 |
H600 |
H750 |
H850 |
H1000 |
H1000 |
|
D780 |
D780 |
D880 |
D880 |
D980 |
D1180 |
D1380 |
|
W810 |
W910 |
W910 |
W1010 |
W1210 |
W1410 |
W1410 |
|
H1500 |
H1500 |
H1600 |
H1750 |
H1850 |
H2000 |
H2000 |
1.1 Model(Examples): THP225
1.2 Temperature range:0℃、-20℃、-30℃、-40℃、-70℃~100℃、120℃、150℃;
1.3 Humidity range:30 ~ 98%R.H;
(refer to temperature&humidity control range drawing)
1.4 Inside capacity: 2225L
1.5 Dimension of working chamber(mm): D500×W600×H750
1.6 External dimensions Probably(mm):D880×W1010×H1750
1.7 Temperature fluctuation:±0.5℃;
1.8 Temperature accuracy:≤±2℃;
1.11 Weight capacity: 325 KG
2. Air Conditioning and Refrigerating system
2.1 the way of air conditioning:compel air to flow, balance to debug temperature
and humidity;
2.2 air circulation:fan;
2.3 the way of heating:open-element Nichrome heater;
2.4 Refrigerating system:Tecumseh, made in France, low back pressure;
2.5 The way of refrigerating:air cooling;
2.6 Refrigeration agent: environment protection type R404A(-40~-70&R23)
2.7 Refrigerating elements: expanding valve、electromagnetism valve、pressure controller、filtration、close valve, etc。
3. Measuring and Controlling system
3.1 Temperature measuring sensor: Pt100;
3.2 Humidity measuring sensor: Pt100;
3.3 Controlling system:Samwontech-TEMI880 ,Programmable Temperature & Humidity Controller;made in Korea.
² Controller: TEMI touch temperature&humidity controller originally imported from Korea, including double channel of temperature, humidity.
² Data display mode: graphic display, numerical display.
² Applicable program capacity: max. 120 groups (1200 segment)
² Control setting precision: temperature 0.1℃, humidity 0.1%RH, time 1S.
² Time setting: 0~99H59min
² Control mode: PID control
² Setting mode: touch type Chinese-English menu operation
² Segment time setting: 0~99H59min
² Temperature & humidity detector: PT100 platinum resistance
² One R232 communication port
3.4 The main electric elements: the most electric units are import;
3.5 Operational fashion: program&constant function。
3.6 For printing:Paperless date recorder.
4. Safety Device
4.1 Protected when the workroom’s temperature overtakes;
4.2 Protected when the blast equipment over loading;
4.3 Protected when add wet system were broke off;
4.4 Protected when compressor over loading;
4.5 Protected when compressor goes beyond pressure;
5. Test Standard
5.1 GB/T2423.1—89
5.2 GB/T2423.4—93 Db
5.3 GJB150.4—86
6. Operation Condition
6.1 Environment temperature:5 ~ 30℃;
6.2 Relative humidity:≤85% R.H
6.3 The air pressure:86 ~ 106 kpa;
6.4 Voltage:380V,50HZ;
7. Chamber
7.1 External frame: A3 OR SUS 304 stainless steel;
7.2 Internal coating: SUS 304 stainless steel;
7.3 Heat preservation material: exceed thin glass cotton
7.4 Multipane inspection window:60D×265W×375H mm
7.5 Cable hole:ф50
7.6 Task lighting:12W;
7.7 Truckless under The bottom of the chamber for moving conveniently
7.8 Filter trap is easy to take apart and maintain